COMMON
MODE VOLTAGE MEASUREMENTS COMPARISON
FOR CISPR 22 CONDUCTED EMISSIONS MEASUREMENTS
Purpose
of Investigation
CISPR 22 (1997) requires the measurement of common mode voltages on
Information Technology Equipment (ITE) cabling to determine conducted
emission levels over the 150 kHz to 30 MHz range. Two techniques that
do not require breaking into the insulation of the cable under test
to make galvanic contact with wires under test are a Capacitive Voltage
Probe and a Capacitve Clamp. This application note compares making common
mode voltage measurements with these two instruments in a simple and
controllable test setup.
CISPR
22 Annex C.1.3 provides the guidance for making this voltage measurement,
and indicates that the common mode voltage measurement is to be made
no more than 80 cm from the Equipment Under Test (EUT). The length of
the cable under test from the voltage measurement location to the Auxiliary/Associated
(AE) can be any length. The effect on the
accuracy of the measured common mode voltage due to uncontrolled impedance/cable
length resulting in a voltage standing wave is addressed in the measurements
shown in this application note.
CISPR
22 requires the common mode capacitive loading to the ITE cable under
test by the voltage measurement technique to be less than 5 pF. Measurements
were made to determine the common mode capacitive loading of the Capacitive
Voltage Probe and the Capacitive Clamp to the cable under test.
Voltage Measurement Techniques
The Capacitive Voltage Probe (CVP) used for these measurements was a
Fischer Custom Communications, Inc. Model F-CVP-1. The cable under test
is placed through the cylindrical aperture of the CVP.
The
Capacitive Clamp (CC) used for these measurements was an EM Test Model
HFK. This is the standard Capacitive Clamp commonly used for injection
of CW and fast transients into cable bundles. To use the CC as a voltage
probe requires the use of a traditional high impedance, low capacitance
voltage probe. The active FET Voltage Probe (FET VP) used in conjunction
with all measurements made with the Capacitive Clamp was a Cal Test
Model CT2865 which is rated at 750 MHz bandwidth with 2 pF of loading.
The test setup used allowed voltage measurements to be made with the
FET VP without the CC.
Test Setup
Test setup is shown in Figure 1. A Hewlett Packard 4396B configured
as a Network Analyzer was used to provide a CW stimulus from a 50 W
source impedance. The test load was a 147 W resistor located at the
end of a 2.7 meter length of #14 wire (insulated) mounted over a 30
cm wide copper ground plate. The 147 W was selected to be consistent
with the 150 W load cited by CISPR 22. The 2.7 meter length was chosen
for two reasons. First, this length allowed standing waves to develop.
Second, the wire length is a conceivable length in a laboratory environment.
Voltage measurement positions denoted A, A, B, C and C as
shown in Figure 1 were used for the measurements in this document.
CVP
Measurements
The voltage on the #14 wire was measured at positions A, B, and C as
shown in Figure 1. The results of the measured voltages at A, B, and
C for this set of measurements are shown in Figure 3.
These
measurements show the effect of the standing wave above a few MHz generated
on the #14 wire due to the mismatch of the 50 W source impedance and
the 147 W load. The difference in measured signal at 30 MHz is about
4 dB from position A to position C.
FET
VP Measurements
The measured voltage at positions A, B, and C as measured
with the FET VP is shown in Figure 3. The FET VP and the CVP measurements
(Figure 3) are about 2 dB apart.
Capacitive
Clamp Measurements
The data in Figure 3 have been corrected only for the FET VP calibration
curve shown in Figure 2. Based on the theory of the CC and FET VP operation
cited above, only the FET VP should affect the voltage measurement.
Capacitive
Loading
To determine how the capacitance of the #14 wire over the ground plane
was loaded by the CVP and the CC, the Hewlett Packard 4275A LCR meter
was used to measure the capacitance for these various physical configurations.
The HP 4275A was placed at one end of the #14 wire (position A in Figure
1). At the other end of the #14 wire, the 147 W load was
removed, and the #14 wire open circuited.
The
CVP (including foam insert) was centered in the middle of the 2 meters
of #14 wire. The total capacitance was measured with the HP 4275A at
100 kHz, 1 MHz, and 10 MHz. The CVP was then removed, keeping the #14
wire in the same physical position. The capacitance was then re-measured.
These
measurements were repeated for the CC, keeping the #14 wire in the same
relative position dictated by the CC size. In addition, the capacitance
measurements were made with and without the FET VP connected to the
CC. The output of the FET VP was loaded with a 50 W termination, and
the low side (shield) of the 50 W termination was connected to the ground
plane.
These
capacitance measurements are shown in the Table below. It can be seen
that the CVP generally adds 3 pF of capacitive loading to the #14 wire.
The Capacitive Clamp without the FET VP adds 25 to 31 pF to the #14
wire. With the FET VP attached, the CC adds 28 to 32 pF to the #14 wire.
The CISPR 22 requirement is for the VP to be used in conjunction with
the Capacitive Clamp to have less than 5 pF loading from the wire/bundle
under test to ground.
|