| This small
shielded loop sensor can be used with a spectrum analyzer or receiver to detect
emissions over a wide frequency spectrum. The sensor is capable of detecting
radiation from currents on surface as well as wires and holes in shields.
The loop area is ideal for resolving changes over small distances, making it useful for diagnostic testing of circuitboards, cabling and shielded structures. The probe has excellent sensitivity and is useful in the "near field". However, since the absolute value of E/H for any given radiation source varies with distance as well as parameters, the output is relative and uncalibrated. In order to provide a degree of calibration, each sensor is calibrated in a TEM cell over the frequency range of 10 MHz - 3 GHz with a 377 ohm impedance. When exposed to a field of 6 amperes/meter the sensor output is typically 282 microvolts at 10 MHz and increases to 11 millivolts at 3 GHz |
||||||||||||||||||||
|
||||||||||||||||||||
|
||||||||||||||||||||
![]() |
||||||||||||||||||||
| ©
2004 Copyright FISCHER CUSTOM COMMUNICATIONS, INC. All rights reserved 20603 Earl Street Torrance, CA 90503 Tel: (310) 303-3300 Fax: (310) 371-6268 E-mail: sales@fischercc.com |
||||||||||||||||||||